ZENEX Multi-System
R-System
Prosthetic
Abutment
Impression Components
Prosthetic Flow Chart
Abutment
Healing Abutment
- Use Healing Abutment fits for the diameter of abutment.
- Select specification fits for fixture connection.
- Tighten with 1.2 Hex Driver by hand
- Recommended tightening torque: 5~8Ncm
- Select appropriate fixture height upon depth of implant placement.
- Select specification fits for fixture connection.
- Tighten with 1.2 Hex Driver by hand
- Recommended tightening torque: 5~8Ncm
- Abutment for manufacturing Screw-retained type temporary prosthesis
- Select specification fits for fixture connection.
- Fixture Level Impression
- Tighten with 1.2 Hex Driver
- Recommended tightening torque: 20Ncm
- To order as a set as 'Abutment + Abutment Screw' :
Product Code + S (ex: TARRH4010S)
- Abutment for manufacturing Cement/Combination-retained type prosthesis
- Select specification fits for fixture connection.
- Customized by grinding (need to be maintained at least 3.0mm of Abutment Length above Fixture Platform)
- Fixture Level Impression
- Tighten with 1.2 Hex Driver
- Recommended tightening torque: 30Ncm
- To order as a set as 'Abutment + Abutment Screw' :
Product Code + S (ex: CARRH40205S)
Angled Abutment
- Abutment for manufacturing Cement/Combination-retained type prosthesis
- Various types of Angle (15º & 25º)
- Select specification fits for fixture connection.
- Fixture Level Impression
- Can be positioned in 12 directions by selecting A or B type
- Tighten with 1.2 Hex Driver
- Recommended tightening torque: 30Ncm
- To order as a set as 'Abutment + Abutment Screw' :
Product Code + S (ex: AARRA40152S)
FreeMilling Abutment
- Abutment for manufacturing Cement/Combination-retained type prosthesis
- Used when creating free marginal space for Abutment
- Select specification fits for fixture connection.
- Customized by grinding (need to be maintained at least 3.0mm of Abutment Length above Fixture Platform)
- Fixture Level Impression
- Tighten with 1.2 Hex Driver
- Recommended tightening torque: 30Ncm
- To order as a set as 'Abutment + Abutment Screw' :
Product Code + S (ex: FMARRH40209S)
CCM Cast Abutment
- Abutment for manufacturing customized abutment in difficult and complicated cases
- Select specification fits for fixture connection.
- Fixture Level Impression
- Casting with non-precious alloy for manufacturing customized prosthesis
- Melting point of CCM: 1,400 ~ 1,550℃
- Tighten with 1.2 Hex Driver
- Recommended tightening torque: 30Ncm
- To order as a set as 'Abutment + Abutment Screw' :
Product Code + S (ex: CCARRH4010S)
Multi Straight Abutment
- Abutment for manufacturing screw-retained prosthesis in Multiple Case
- Same platform as Multi Angled Abutment
- Move into internal oral part by using exclusive Abutment Carrier (Code: MSACR48)
- Tighten with exclusive driver (Code: MSADSR20)
- Recommended tightening torque: 30Ncm
- To order as a set as 'Abutment + Carrier' :
Product Code + S (ex: MSARR4815S)
Multi Angled Abutment
- Abutment for manufacturing screw-retained prosthesis in Multiple Case
- Abutment of various angles (17º, 30º) for various angled of implant insertion path
- Same platform as Multi Straight Abutment
- Compensation of fixture placement angle up to 108º
- Connect by using exclusive Abutment Carrier (Code: MAACRMC)
- Tighten with 1.2 Hex driver
- Recommended tightening torque: 30Ncm
- Angled Abutment Screw (MAASRR20) included
- To order as a set as 'Abutment + Carrier':
Product Code + S (ex: MAARRH481725S)
- Three kinds of cylinder component for Multi Abutments
- Multi Healing Cap
- Abutment for overdenture using O-ring attachment
- Compensation of mounting angle up to 20º
- Tighten with exclusive Ball Abutment Driver (Code: BAD24)
- Recommended tightening torque: 30Ncm
- Ball Abutment Cap Set & Ball Abutment Retainer Set
- 5 ea of O-ring are composed with a Ball Abutment O-ring Set
- Compensation of fixture placement up to 40º
- 1.5mm lower vertifical dimension
- Composition of multiple attachments with stable retention
- Connect with exclusive driver
- Recommended tightening torque: 30Ncm
Kerator Angled Abutment
- 1.5mm lower vertifical dimension
- Composition of multiple attachments with stable retention
- Connect with 1.2 Hex driver
- Recommended tightening torque: 30Ncm
- CDPH Set: White spacer / Denture Cap & Three kinds of Male Cap (Red, Blue, Pink)
- Seven kinds of Male Cap, according to the retention
Impression Components
Pick-up Impression Coping
- Components for Fixture Level Impression
- Open Tray Type
- Select specification fits for fixture connection
- Enables accurate impression with design of stably fixed in impression material
- Tighten with 1.2 Hex Driver by hand
- To order as a set as 'Impression Coping Body + Guide Pin' :
[For 11mm Length] Product Code + S (ex: PIRRH40SS)
[For 15mm Length] Product Code + S (ex: PIRRH40LS)
Transfer Impression Coping
- Components for Fixture Level Impression
- Closed Tray Type
- Select specification fits for fixture connection
- Streamlined Shape: Easy to transfer
- Anti-Rotation Grooves accord with Hex of fixture
- Tighten with 1.2 Hex Driver by hand
- To order as a set as 'Impression Coping Body + Guide Pin' :
[For 9mm Length] Product Code + S (ex: TIRRH40SS)
[For 13mm Length] Product Code + S (ex: TIRRH40LS)
Fixture Lab Analog
- Lab Analog for Fixture Level Impression
- Select among Fixture Diameter (Ø 3.5 / Ø 4.0 and above)
Multi Impression Coping
- Two kinds of impression coping for taking Abutment Level impression for Multi Abutments
- Open Tray Type for Multi Pick-up Impression Coping
- Closed Tray Type for Multi Transfer Impression Coping
- Tighten with 1.2 Hex Driver by hand
- In the Multi Pick-up Impression Coping, Multi Pick-up Impression Coping Guide Pin (MPICSRGP) included.
- To order a Multi Pick-up Impression Coping as a set
as 'Impression Coping Body + Guide Pin':
Product Code + S (ex: MPICSR48S)
Multi Lab Analog
- Lab Analog for Multi Abutment
Ball Abutment Lab Analog
- Lab Analog for Ball Abutment
Kerator Impression Coping
- Pick-up Impression coping for Kerator Abutment
- Closed Tray Type
Kerator Lab Analog
- Lab Analog for Kerator Abutment
Digital
ScanBody
- ScanBody for manufacturing customized Titanium abutment
- Use for Oral scan (Model scan available as well)
- Tighten with 1.2 Hex Driver by hand
- To order as a set as 'ScanBody + ScanBody Screw' :
Product Code + S (ex: RSBR4310S)
Ti-Link Abutment
- Abutment for manufacturing Cement/Combination-retained type prosthesis
- For manufacturing custom abutment (Titanium & Zirconia) and crown by CAD/CAM equipment
- Select specification fits for fixture Connection
- Use exclusive library for Zenex Multi Implant system
- Fixture Level Impression
- Tighten with 1.2 Hex driver
- Recommended tightening torque: 30Ncm
- To order as a set as 'Abutment + Screw' :
Product Code + S (ex: TLARRH4010S)
Ti-Blank Abutment
- Manufacturing customized abutment with milling machine
- Select specification fits for fixture Connection
- Digital Impression
- Tighten with 1.2 Hex driver
- Recommended tightening torque: 30Ncm
- Product line-up applied for various milling machine brands
(Milling machine manufacturer: Doowon, Manix, Vatech, RND) - To order as a set as 'Abutment + Screw' :
Product Code + S (ex: TBARRH10AS)
Multi Ti-Link Cylinder
- Multi Cylinder for making Combination-retained prosthesis from Multi Abutment.
- Use when producing Customized Abutment (Titanium & Zirconia) and Crown by CAD/CAM equipment.
- Use exclusive implant library of Zenex Multi Implant System
- Abutment Level Impression
- Tighten with 1.2 Hex driver
- Recommended tightening torque: 20Ncm
- Multi Cylinder Screw (MTCSR23) included
- To order as a set as 'Cylinder + Multi Cylinder Screw' :
Product Code + S (ex: MTLCSRN48S)